VITROS CHEMISTRY PRODUCTS PHYT SLIDES 8298671

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-12-19 for VITROS CHEMISTRY PRODUCTS PHYT SLIDES 8298671 manufactured by Ortho-clinical Diagnostics.

Event Text Entries

[131911876] The investigation determined that higher than expected vitros phenytoin (phyt) results were obtained on a patient sample and a quality control (qc) fluid and a higher than expected vitros carbamazepine (crbm) result was obtained from a patient sample when tested using vitros chemistry products phyt slides and vitros chemistry products crbm slides in combination with two different vitros 5600 integrated systems. The most likely assignable cause of the event was a sample related issue. The affected patient correlation samples appeared to be slightly viscous, although no sample metering condition codes were generated when the samples were processed. In addition, some samples may have been stored refrigerated for up to 2 weeks. However, the samples used for the vitros phyt correlation were plasma samples stored refrigerated on average for 7 days, while the vitros crbm correlation were plasma samples stored refrigerated on average for 5 days. Per the vitros phyt instructions for use (ifu), samples can be stored refrigerated for =7 days, while per the vitros crbm ifu, samples can be stored refrigerated for =5 days. Improper sample handling could not be ruled out as contributing to the event as no information was provided concerning the centrifugation procedure or if the samples were centrifuged a second time prior to the retesting events. It is possible that cellular debris, due to poor sample preparation, was present in the affected samples, although this could not be confirmed. A review of vitros phyt quality control results processed on (b)(4) did show some imprecision within the timeframe of the event. Therefore, an instrument or slide related issue cannot be completely ruled out for the higher than expected vitros phyt results. Ongoing tracking and trending of complaints has not identified any signals that would suggest there is a systemic issue with either vitros phyt reagent lot 2617-0168-2514 or vitros crbm reagent lot 3942-0102-0838. The customer later processed four alternate patient samples and both biorad qc levels on analyzer (b)(4) using both lots of immunowash fluid (iwf) and vitros phyt slides. They also ran an additional four patient samples and both biorad qc levels on analyzer (b)(4) using both lots of iwf and vitros crbm slides. All correlation patient results were deemed acceptable.
Patient Sequence No: 1, Text Type: N, H10


[131911877] A customer obtained higher than expected results on patient samples used for correlation testing between a previous and new lot of immunowash fluid (iwf). The patient samples were tested using vitros chemistry products phenytoin (phyt) slides and vitros chemistry products carbamazepine (crbm) slides on two different vitros 5600 integrated systems. A higher than expected quality control fluid result was discovered later. Vitros phyt slide lot 2617-0168-2514 tested on (b)(4). Patient sample a vitros phyt result of 42. 8 umol/l vs. The expected result of 29. 6 umol/l. Vitros phyt slide lot 2617-0168-2514 tested on (b)(4). Non-vitros biorad immunoassay plus control lot 27730 l3 vitros phyt result of 147. 5 umol/l vs. The expected result of 113. 0 umol/l. Vitros crbm slide lot 3942-0102-0838 tested on (b)(4). Patient sample q vitros crbm result of 64. 8 umol/l vs. The expected result of 44. 3 umol/l. Biased results of the magnitude and direction observed may lead to inappropriate physician action if they were to occur undetected on patient samples. The vitros phyt and crbm results were not reported outside the lab. There were no allegations of patient harm as a result of this event. This report is number 1 of 3 mdr? S for this event. Three (3) 3500a forms are being submitted for this event as 3 devices were involved. This report corresponds to ortho clinical diagnostics inc. Complaint numbers (b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1319809-2018-00166
MDR Report Key8181031
Date Received2018-12-19
Date of Report2018-12-19
Date of Event2018-10-01
Date Mfgr Received2018-11-23
Device Manufacturer Date2018-08-25
Date Added to Maude2018-12-19
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. JAMES A STEVENS
Manufacturer Street100 INDIGO CREEK DRIVE
Manufacturer CityROCHESTER NY 14626
Manufacturer CountryUS
Manufacturer Postal14626
Manufacturer Phone5854533000
Manufacturer G1ORTHO-CLINICAL DIAGNOSTICS
Manufacturer Street513 TECHNOLOGY BLVD.
Manufacturer CityROCHESTER NY 14652
Manufacturer CountryUS
Manufacturer Postal Code14652
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameVITROS CHEMISTRY PRODUCTS PHYT SLIDES
Generic NameIN VITRO DIAGNOSTICS
Product CodeDIP
Date Received2018-12-19
Catalog Number8298671
Lot Number2617-0168-2514
Device Expiration Date2019-01-15
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerORTHO-CLINICAL DIAGNOSTICS
Manufacturer Address100 INDIGO CREEK DRIVE ROCHESTER NY 14626 US 14626


Patients

Patient NumberTreatmentOutcomeDate
10 2018-12-19

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.