TUR IRRIGATION SET ARC4005P

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2018-12-19 for TUR IRRIGATION SET ARC4005P manufactured by Baxter Healthcare - Cali.

Event Text Entries

[130918758] The device has been received and the evaluation is in progress. Should additional relevant information become available, a supplemental report will be submitted.
Patient Sequence No: 1, Text Type: N, H10


[130918759] It was reported that the clamp of a tur irrigation set was missing. The missing clamp? Was identified before to patients use. There was no patient involvement. No additional information is available.
Patient Sequence No: 1, Text Type: D, B5


[133371258] The device was received for evaluation. A visual inspection was made and it was observed that the sample was correctly assembled and packed according to local specifications. Additionally, the lower clamp was absent. The reported condition was verified. The cause of the condition was related to human production during manufacturing. A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot. A nonconformance has been opened to address this issue. Should additional relevant information become available, a supplemental report will be submitted.
Patient Sequence No: 1, Text Type: N, H10


[133371259]
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1416980-2018-07906
MDR Report Key8181552
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2018-12-19
Date of Report2019-01-16
Date Mfgr Received2018-12-22
Device Manufacturer Date2018-09-24
Date Added to Maude2018-12-19
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationPHARMACIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer Street25212 W. ILLINOIS ROUTE 120
Manufacturer CityROUND LAKE IL 60073
Manufacturer CountryUS
Manufacturer Postal60073
Manufacturer Phone2242702068
Manufacturer G1BAXTER HEALTHCARE - CALI
Manufacturer StreetCALLE 36 NO. 2C-22 APARTADO AERO 2446
Manufacturer CityCALI
Manufacturer CountryCO
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTUR IRRIGATION SET
Generic NameSYSTEM, IRRIGATION, UROLOGICAL
Product CodeLJH
Date Received2018-12-19
Returned To Mfg2018-12-06
Model NumberNA
Catalog NumberARC4005P
Lot NumberSE18IM1
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBAXTER HEALTHCARE - CALI
Manufacturer AddressCALI


Patients

Patient NumberTreatmentOutcomeDate
10 2018-12-19

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.