[130985937] Neonatal transport isolette/oxygen delivery-blender device failed during (and prior to) neonatal intensive care unit (nicu) transport. This caused inability to bringing patient back on a low fio2/low flow nasal cannula, which was current therapy ordered. History of event: blender alarm noted upon initial assessment around 10:00 am-prior to leaving the facility on transport. Three additional rts evaluated the blender alarm and came to the conclusion that biomed needed to be involved. Biomed was not immediately available, so engineering came immediately and then was able to involve biomed after further assessment. One of the air regulators was found to have a large leak no matter what troubleshooting was completed.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number8183858
MDR Report Key8183858
Date Received2018-12-20
Date of Report2018-12-18
Date of Event2018-12-06
Report Date2018-12-19
Date Reported to FDA2018-12-19
Date Reported to Mfgr2018-12-20
Date Added to Maude2018-12-20
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVOYAGER TRANSPORT INCUBATOR
Generic NameINCUBATOR, NEONATAL TRANSPORT
Product CodeFPL
Date Received2018-12-20
Catalog Number20035655
Device AvailabilityY
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerINTERNATIONAL BIOMEDICAL, LTD.
Manufacturer Address8206 CROSS PARK DR AUSTIN TX 78754 US 78754


Patients

Patient NumberTreatmentOutcomeDate
10 2018-12-20