VITAL SIGNS? FETAL SPIRAL ELECTRODE FOR COROMETRICS 7000AAO

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2018-12-20 for VITAL SIGNS? FETAL SPIRAL ELECTRODE FOR COROMETRICS 7000AAO manufactured by Vyaire Medical.

Event Text Entries

[131065602] Vyaire complaint (b)(4). Any additional information provided by the customer will be included in a follow-up report. Due diligence is ongoing. At this time, vyaire is pursuing additional information and sample availability for evaluation.
Patient Sequence No: 1, Text Type: N, H10


[131065603] A customer contacted vyaire medical to report that two wires (pink & blue) seemed to be twisted in the wrong direction. When the end user was attempting to pull the wires apart to disengage from the (unborn) baby's head, instead of the small corkscrew wire that is at the end of the electrode unscrewing, it was screwing deeper into the unborn baby's head. This is the first time the end user customer experienced the issue. There was no indication of patient harm or medical intervention during initial contact.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3003678543-2018-00001
MDR Report Key8184893
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2018-12-20
Date of Report2018-11-30
Date of Event2018-11-30
Date Mfgr Received2018-11-30
Date Added to Maude2018-12-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. STANLEY TAN
Manufacturer Street22745 SAVI RANCH PKWY
Manufacturer CityYORBA LINDA CA 92887
Manufacturer CountryUS
Manufacturer Postal92887
Manufacturer Phone7149193324
Manufacturer G1PRECISION CONCEPTS COSTA RICA SA
Manufacturer StreetSARET FREE ZONE BUILDING D3 P.O. BOX 74-4002
Manufacturer CityALAJUELA,
Manufacturer CountryCS
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVITAL SIGNS? FETAL SPIRAL ELECTRODE FOR COROMETRICS
Generic NameELECTRODE, CIRCULAR (SPIRAL), SCALP AND APPLICATOR
Product CodeHGP
Date Received2018-12-20
Catalog Number7000AAO
Lot NumberS07181-F, S35171-F
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerVYAIRE MEDICAL
Manufacturer Address22745 SAVI RANCH PKWY YORBA LINDA CA 92887 US 92887


Patients

Patient NumberTreatmentOutcomeDate
10 2018-12-20

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.