5487

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2018-12-20 for 5487 manufactured by Medtronic, Inc..

Event Text Entries

[131065349] If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[131065350] It was reported that the patient cable was being used in conjunction with an external pulse generator (epg) on a patient post-operative. The patient? S cardiac rhythm was driven by the epg system when there was a cessation of stimulation. The patient cable was replaced, and the issue was resolved. The status of the original patient cable is unknown. No patient complications have been reported as a result of this event.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2182208-2018-02306
MDR Report Key8186053
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2018-12-20
Date of Report2019-03-04
Date of Event2018-11-28
Date Mfgr Received2019-02-27
Date Added to Maude2018-12-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactLISA ROBERTSON
Manufacturer Street8200 CORAL SEA ST NE
Manufacturer CityMOUNDS VIEW MN 55112
Manufacturer CountryUS
Manufacturer Postal55112
Manufacturer Phone7635262723
Manufacturer G1MEDTRONIC, INC.
Manufacturer Street8200 CORAL SEA STREET NE
Manufacturer CityMOUNDS VIEW MN 55112
Manufacturer CountryUS
Manufacturer Postal Code55112
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameNA
Generic NameCABLE, TRANSDUCER AND ELECTRODE, PATIENT, (INCLUDING CONNECTOR)
Product CodeDSA
Date Received2018-12-20
Returned To Mfg2019-01-29
Model Number5487
Catalog Number5487
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMEDTRONIC, INC.
Manufacturer Address8200 CORAL SEA STREET NE MOUNDS VIEW MN 55112 US 55112


Patients

Patient NumberTreatmentOutcomeDate
101. Life Threatening; 2. Required No Informationntervention 2018-12-20