ULTRA-CATCH NT (1.8 FR X 115 CM) STONE RETRIEVAL DEVICE NT4W18115

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-12-20 for ULTRA-CATCH NT (1.8 FR X 115 CM) STONE RETRIEVAL DEVICE NT4W18115 manufactured by Gyrus Acmi, Inc.

Event Text Entries

[131493366] The device was returned to olympus medical systems corp in (b)(4) for evaluation. A visual inspection of the device found a piece of the basket wires broken. The end of the broken wire appeared to have been melted. The basket does not deploy and retract smoothly and the sheath of the device was kinked. A long clear colored ribbon-like material was returned with the device. The foreign material does not appear to be a component of the device. The cause of the melted basket and the resistance the user's experienced is unknown. The olympus sales representative further reported that during the middle of the procedure, the basket was deployed inside the patient without issue. There was no unusual behavior from the device or equipment prior to the reported event. The intended procedure was completed. The cause of the reported event cannot be confirmed, however, olympus will continue to investigate. If additional information becomes available, this report will be supplemented accordingly.
Patient Sequence No: 1, Text Type: N, H10


[131493367] Olympus was informed that during a transurethral lithotripsy procedure, the physician felt resistance when the device is being retracted. The tip of the device was then inspected subsequent to the procedure, and a coating like material appeared to have been peeled off. There was no patient injury reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2951238-2018-00775
MDR Report Key8187882
Date Received2018-12-20
Date of Report2018-12-20
Date of Event2018-10-15
Date Mfgr Received2018-11-20
Date Added to Maude2018-12-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. CONNIE TUBERA
Manufacturer Street2400 RINGWOOD AVENUE
Manufacturer CitySAN JOSE CA 95131
Manufacturer CountryUS
Manufacturer Postal95131
Manufacturer Phone4089355124
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameULTRA-CATCH NT (1.8 FR X 115 CM) STONE RETRIEVAL DEVICE
Generic NameSTONE RETRIEVAL BASKET
Product CodeFGO
Date Received2018-12-20
Returned To Mfg2018-10-24
Model NumberNT4W18115
Lot NumberMK769245
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerGYRUS ACMI, INC
Manufacturer Address136 TURNPIKE ROAD SOUTHBOROUGH MA 01772 US 01772


Patients

Patient NumberTreatmentOutcomeDate
10 2018-12-20

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.