FUSE FULL SPECTRUM ENDOSCOPY SYSTEM FSC330-C6337 60093

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2018-12-26 for FUSE FULL SPECTRUM ENDOSCOPY SYSTEM FSC330-C6337 60093 manufactured by Boston Scientific Corporation.

Event Text Entries

[131476856] (b)(4). A fuse 1c colonoscope was returned for analysis. A functional evaluation was performed and the reported image loss was confirmed due to a failed charged couple device (ccd). The ccd was replaced to resolve the issue. The observed damage to the returned device is consistent with component failure (expected or random). Therefore, a review and analysis of all available information indicated that the most probable root cause is cause traced to component failure. A review of the device history record (dhr) confirmed that the device met all material, assembly, and product specifications.
Patient Sequence No: 1, Text Type: N, H10


[131476857] It was reported to boston scientific corporation that a fuse 1c colonoscope was used during a colonoscopy procedure performed on (b)(6) 2018. According to the complainant, during the procedure, the video looked blue. The anatomy of the patient was not visible when the video turned blue. Another fuse 1c colonoscope was used to complete the procedure. There were no patient complications reported as a result of this event.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005099803-2018-62247
MDR Report Key8196952
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2018-12-26
Date of Report2018-12-26
Date of Event2018-11-28
Date Mfgr Received2018-11-28
Device Manufacturer Date2015-12-04
Date Added to Maude2018-12-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactCAROLE MORLEY
Manufacturer Street300 BOSTON SCIENTIFIC WAY
Manufacturer CityMARLBOROUGH MA 01752
Manufacturer CountryUS
Manufacturer Postal01752
Manufacturer Phone5086834015
Manufacturer G1ENDOCHOICE, INC.
Manufacturer Street11810 WILLS ROAD
Manufacturer CityALPHARETTA GA 30009
Manufacturer CountryUS
Manufacturer Postal Code30009
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameFUSE FULL SPECTRUM ENDOSCOPY SYSTEM
Generic NameENDOSCOPE, ACCESSORIES, IMAGE POST-PROCESSING FOR COLOR ENHANCEMENT
Product CodePEA
Date Received2018-12-26
Returned To Mfg2018-12-13
Model NumberFSC330-C6337
Catalog Number60093
Lot Number0001000393
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBOSTON SCIENTIFIC CORPORATION
Manufacturer Address300 BOSTON SCIENTIFIC WAY MARLBOROUGH MA 01752 US 01752


Patients

Patient NumberTreatmentOutcomeDate
10 2018-12-26

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.