1.5 LACTOSORB SYSTEM 1.5 X 4 MM LACTOSORB SCREW N/A 915-2315

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a distributor,foreign,health pr report with the FDA on 2018-12-27 for 1.5 LACTOSORB SYSTEM 1.5 X 4 MM LACTOSORB SCREW N/A 915-2315 manufactured by Biomet Microfixation.

Event Text Entries

[131523350] (b)(4). Foreign country: (b)(6). Customer has indicated that the product will not be returned to zimmer biomet for investigation, it was discarded. The investigation is in process. Once the investigation has been completed, a follow-up mdr will be submitted.
Patient Sequence No: 1, Text Type: N, H10


[131523351] It was reported that during an open reduction procedure the hex screw head of the first screw was wrenched off before it was inserted into the patient's bone completely. The surgeon could not fix a second screw due to the idling of that screw. The surgery was completed using an emergency screw. Attempts have been made and no further information has been provided. No additional patient consequences were reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number0001032347-2018-00896
MDR Report Key8199786
Report SourceDISTRIBUTOR,FOREIGN,HEALTH PR
Date Received2018-12-27
Date of Report2018-12-27
Date of Event2018-08-10
Date Mfgr Received2018-11-29
Device Manufacturer Date2018-02-13
Date Added to Maude2018-12-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS. MICHELLE COLE
Manufacturer Street1520 TRADEPORT DRIVE
Manufacturer CityJACKSONVILLE FL 32218
Manufacturer CountryUS
Manufacturer Postal32218
Manufacturer Phone9047414400
Manufacturer G1BIOMET MICROFIXATION
Manufacturer Street1520 TRADEPORT DRIVE
Manufacturer CityJACKSONVILLE FL 32218
Manufacturer CountryUS
Manufacturer Postal Code32218
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand Name1.5 LACTOSORB SYSTEM 1.5 X 4 MM LACTOSORB SCREW
Generic NamePOLYMER, EAR, NOSE, AND THROAT, SYNTHETIC, ABSORBABLE
Product CodeNHB
Date Received2018-12-27
Model NumberN/A
Catalog Number915-2315
Lot Number960000
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBIOMET MICROFIXATION
Manufacturer Address1520 TRADEPORT DRIVE JACKSONVILLE FL 32218 US 32218


Patients

Patient NumberTreatmentOutcomeDate
10 2018-12-27

© 2026 FDA.report