P500 MRS CONTROL UNIT P005787RENT02

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-12-27 for P500 MRS CONTROL UNIT P005787RENT02 manufactured by Hill-rom Batesville.

Event Text Entries

[131743900] The hill-rom technician has picked up the unit for evaluation and repair. It is necessary for the p500 to have an effective maintenance program. We recommend that you do annual preventive maintenance. Control unit; make sure of these: the out of bed sensors operate correctly. Mattress assembly; make sure of these: the out of bed sensors and head of bed sensor mounting are not damaged. Repair or replace as necessary. A search of the hill-rom maintenance records showed hill-rom performed preventative maintenance on this unit in (b)(6) 2018. It is unknown if the facility performed any other preventative maintenance on this unit. No further information is available on the repair of the bed at this time. The investigation is ongoing, however if any additional relevant information is identified following completion of the investigation, the additional relevant information will be submitted in a supplemental report.
Patient Sequence No: 1, Text Type: N, H10


[131743901] Hill-rom received a report from the account stating the bed exit alarm was not working. The bed was located at the account. There was no patient/user injury reported. This report was filed in our complaint handling system as (b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1824206-2018-00497
MDR Report Key8200215
Date Received2018-12-27
Date of Report2018-11-30
Date of Event2018-11-30
Date Mfgr Received2018-11-30
Device Manufacturer Date2010-08-12
Date Added to Maude2018-12-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactEMILY MITCHELL
Manufacturer Street1069 STATE ROUTE 46 EAST
Manufacturer CityBATESVILLE IN 47006
Manufacturer CountryUS
Manufacturer Postal47006
Manufacturer Phone8129310130
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameP500 MRS CONTROL UNIT
Generic NameBED, FLOTATION THERAPY, POWERED
Product CodeIOQ
Date Received2018-12-27
Model NumberP005787RENT02
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerHILL-ROM BATESVILLE
Manufacturer Address1069 STATE ROUTE 46 EAST BATESVILLE IN 47006 US 47006


Patients

Patient NumberTreatmentOutcomeDate
10 2018-12-27

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.