INDEPENDENCE IBOT 4000 MOBILITY SYSTEM IT004061

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 04 report with the FDA on 2007-02-09 for INDEPENDENCE IBOT 4000 MOBILITY SYSTEM IT004061 manufactured by Independence Technology, L.l.c..

Event Text Entries

[20989586] User called to request service stating that he tipped the device over twice while on vacation in europe. For 1st event, user reported that while descending a set of stairs with his wife's assistance, she pulled back and the device fell back on her. User would not discuss if his wife was injured - stated he did not care if information regarding injury was required for the investigation. For 2nd event, user reported that he has just finished climbing stairs and transitioned to balance function. User states the device fell over backwards. The user was not injured. While no injury was confirmed as a result of these events, this mdr is being filed due to the absence of confirmation of injury to the users wife, when the device reportedly fell back on her. This report corresponds to independence technology complaint # 782504 and 782793.
Patient Sequence No: 1, Text Type: D, B5


[21064668] Service was dispatched to clear a service wrench in response to a call from a user. A report on field service activity (sar) and a device checkout record (fcr) was forwarded to the complaint handling unit (chu) per standard operating procedure. The user has not reported any recurrence of the described event. The ecf (electronic configuration file) was evaluated once returned to the mfg. For the 1st event, the ecf shows there is only 1 controller failure in the file, which occurred in 2006, at 07:28 gmt. This was for a pitch limit exceeded, which indicates the assistant pulled too far back on the device while descending the stairs. It is likely, while descending, the assistant accidentally pulled back too much when slowing down the rotation, and caused the device to climb back up one step, and her footing may have not been prepared for that. As the file was not retrieved until one month later, only this single alarm is available. In reference to the 2 events in the order stated by the consumer, the stair event could not have caused a service wrench, since the 2nd event mentions transition to balance, which would be disable if the first event caused a wrench. For the 2nd event, no other alarms occurred in 2006, prior to the pitch limit controller failure. Event data was not available. No device malfunction is indicated by the ecf review.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3003508375-2007-00001
MDR Report Key820140
Report Source04
Date Received2007-02-09
Date of Report2007-02-07
Date of Event2007-01-09
Device Manufacturer Date2005-09-01
Date Added to Maude2007-02-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer ContactMICHAEL O'MEARA, DIRECTOR
Manufacturer Street45 TECHNOLOGY DRIVE
Manufacturer CityWARREN NJ 07059
Manufacturer CountryUS
Manufacturer Postal07059
Manufacturer Phone9084122200
Manufacturer G1CREATIVE TECHNOLOGY SERVICES
Manufacturer Street7444 HAGGERTY ROAD
Manufacturer CityCANTON MI 48187
Manufacturer CountryUS
Manufacturer Postal Code48187
Single Use3
Remedial ActionRP
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameINDEPENDENCE IBOT 4000 MOBILITY SYSTEM
Generic NameSTAIR CLIMBING WHEELCHAIR
Product CodeIMK
Date Received2007-02-09
Model NumberNA
Catalog NumberIT004061
Lot Number*
ID Number*
OperatorLAY USER/PATIENT
Device AvailabilityN
Device Eval'ed by MfgrY
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key807610
ManufacturerINDEPENDENCE TECHNOLOGY, L.L.C.
Manufacturer Address45 TECHNOLOGY DR. WARREN NJ 07059 US
Baseline Brand NameINDEPENDENCE IBOT 4000 MOBILITY SYSTEM
Baseline Generic NameSTAIR CLIMBING WHEELCHAIR
Baseline Model NoNA
Baseline Catalog NoIT004061
Baseline ID*


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2007-02-09

© 2024 FDA.report