TONSIL WIRE (SNARE)

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 1993-08-11 for TONSIL WIRE (SNARE) manufactured by Storz.

Event Text Entries

[18197263] End of the tonsil wire broken while removing the right tonsil. Wire piece could not be found in the patient's throat, no clinical problems noted while patient was in or or pacu. Device not labeled for single use. Patient medical status prior to event: satisfactory condition. There was not multiple patient involvement. Invalid data - on device service/maintenance. No data - regarding date last serviced. Service provided by: invalid data. Invalid data - service records availability. No imminent hazard to public health claimed. Invalid data - whether device used as labeled/intended. Device was evaluated after the event. Method of evaluation: actual device involved in incident was evaluated, visual examination. Results of evaluation: component failure, telemetry failure, none or unknown, none or unknown. Conclusion: device failure directly caused event. Certainty of device as cause of or contributor to event: yes. Corrective actions: device discarded. The device was destroyed/disposed of.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number8202
MDR Report Key8202
Date Received1993-08-11
Date of Report1993-07-07
Date of Event1993-06-29
Date Facility Aware1993-06-29
Report Date1993-07-07
Date Reported to FDA1993-07-07
Date Reported to Mfgr1993-07-07
Date Added to Maude1994-07-28
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional0
Initial Report to FDA0
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Generic NameTONSIL WIRE (SNARE)
Product CodeKBZ
Date Received1993-08-11
Device AvailabilityN
Implant FlagN
Device Sequence No1
Device Event Key7877
ManufacturerSTORZ


Patients

Patient NumberTreatmentOutcomeDate
101. Other 1993-08-11

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.