FREE PSA 03289788190

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2019-01-02 for FREE PSA 03289788190 manufactured by Roche Diagnostics.

Event Text Entries

[131786541] This event occurred in (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[131786542] The customer complained of discrepant results for 1 patient sample tested for elecsys total psa immunoassay (total psa) and elecsys free psa immunoassay (free psa) on a cobas e 411 immunoassay analyzer. This medwatch will cover free psa. Refer to medwatch with patient identifier (b)(6) for information on the total psa results. The initial total psa result was 0. 32 ng/ml. The initial free psa result was 0. 444 ng/ml. The repeat total psa result was 0. 297 ng/ml. The repeat free psa result was 0. 430 ng/ml. All results were reported outside of the laboratory. There was no allegation that an adverse event occurred. The e411 analyzer serial number was (b)(4). The customer suspects an interfering factor affecting the results for this patient. The sample was requested for investigation but could not be provided. Based on the information provided, the investigation did not identify a product problem. The cause of the event could not be determined.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2019-00003
MDR Report Key8208104
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2019-01-02
Date of Report2019-01-02
Date of Event2018-12-18
Date Mfgr Received2018-12-18
Date Added to Maude2019-01-02
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameFREE PSA
Generic NameTEST,PROSTATE SPECIFIC ANTIGEN,FREE,(NONCOMPLEXED) TO DISTINGUISH PROSTATE CA
Product CodeMTG
Date Received2019-01-02
Model NumberNA
Catalog Number03289788190
Lot Number330814
Device Expiration Date2019-09-30
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2019-01-02

© 2026 FDA.report