CLINICAL CHEMISTRY LACTATE DEHYDROGENASE 02P56-21

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2019-01-07 for CLINICAL CHEMISTRY LACTATE DEHYDROGENASE 02P56-21 manufactured by Abbott Manufacturing Inc.

Event Text Entries

[133194719] An evaluation is in process. A follow-up report will be submitted when the evaluation is complete. There is no further patient information provided due to privacy issues.
Patient Sequence No: 1, Text Type: N, H10


[133194720] The customer reported falsely elevated ldh results generated on the architect analyzer for one patient. The results provided were: (b)(6) 2018 = 1020u/l which was reported and questioned by the nurse; the sample was retested = 388u/l; another sample from the same patient same day was tested = 201u/l. There was no reported impact to patient management.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1628664-2019-00009
MDR Report Key8223214
Report SourceHEALTH PROFESSIONAL
Date Received2019-01-07
Date of Report2019-01-21
Date of Event2018-12-14
Date Mfgr Received2019-01-08
Date Added to Maude2019-01-07
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNOEMI ROMERO-KONDOS, RN BSN
Manufacturer Street100 ABBOTT PARK ROAD DEPT. 09B9, LCCP1-3
Manufacturer CityABBOTT PARK IL 600643537
Manufacturer CountryUS
Manufacturer Postal600643537
Manufacturer Phone224667-512
Manufacturer G1ABBOTT MANUFACTURING INC
Manufacturer Street1921 HURD DRIVE
Manufacturer CityIRVING TX 75038
Manufacturer CountryUS
Manufacturer Postal Code75038
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCLINICAL CHEMISTRY LACTATE DEHYDROGENASE
Generic NameLACTATE DEHYDROGENASE
Product CodeCFJ
Date Received2019-01-07
Catalog Number02P56-21
Lot Number86459UN18
Device Expiration Date2019-06-11
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerABBOTT MANUFACTURING INC
Manufacturer Address1921 HURD DRIVE IRVING TX 75038 US 75038


Patients

Patient NumberTreatmentOutcomeDate
10 2019-01-07

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.