TORQUE-SHANK, T7X80 CANNULATED, AO CONNECTOR 54098-80

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-01-09 for TORQUE-SHANK, T7X80 CANNULATED, AO CONNECTOR 54098-80 manufactured by I.t.s. Gmbh.

Event Text Entries

[132810585] The device history record (dhr) of the last three shipments of the screwdriver shank 54098-80 and the resp. Screws that were send to general care were inspected and showed no deviations. The quality forms met all dimensional and visual criteria at the time of release with no issues documented during the manufacturing that would contribute to this complaint condition. A review of the raw material device history record revealed no deviation. The material was determined to be conforming and was used as is per product development approval. We have tested the shank with multiple hcs screws from our stock and the shank engaged smoothly in the torque-profile. No deviation resp. Non-conformance detected.
Patient Sequence No: 1, Text Type: N, H10


[132810586] It was reported that a headless compression screw was mismatching with the torque shank. The incident didn't happen in surgery.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3004369035-2016-00007
MDR Report Key8230403
Date Received2019-01-09
Date of Report2019-01-09
Date Mfgr Received2016-05-11
Date Added to Maude2019-01-09
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. FLORIAN GRILL
Manufacturer StreetAUTAL 28
Manufacturer CityLASSNITZH[?]HE, STYRIA 8301
Manufacturer CountryAU
Manufacturer Postal8301
Manufacturer G1I.T.S. GMBH
Manufacturer StreetAUTAL 28
Manufacturer CityLASSNITZH[?]HE, STYRIA 8301
Manufacturer CountryAU
Manufacturer Postal Code8301
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameTORQUE-SHANK, T7X80 CANNULATED, AO CONNECTOR
Generic NameTORQUE-SHANK, T7X80 CANNULATED, AO CONNECTOR
Product CodeLRZ
Date Received2019-01-09
Model Number54098-80
Catalog Number54098-80
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerI.T.S. GMBH
Manufacturer AddressAUTAL 28 LASSNITZH[?]HE, STYRIA 8301 AU 8301


Patients

Patient NumberTreatmentOutcomeDate
10 2019-01-09

© 2026 FDA.report