SMART STAPES PISTON .6 X 4.50 MM 70145928

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,user facility report with the FDA on 2019-01-10 for SMART STAPES PISTON .6 X 4.50 MM 70145928 manufactured by Gyrus Acmi, Inc.

Event Text Entries

[132645779] The device was not returned to olympus for evaluation. The cause of the reported event cannot be determined at this time. However, the instruction for use warns users in the "precautions": care should be taken not to damage the bony structures, surrounding mucosa, chorda tympani, facial nerve, or any external tissue when applying heat to the wire loops. Olympus will continue to investigate this complaint to obtain more detailed information regarding the reported complaint. If additional information becomes available, this report will be updated and supplemented accordingly.
Patient Sequence No: 1, Text Type: N, H10


[132645780] Olympus was informed that during an ear implant procedure, the piston was heated several times due to not crimping and resulted in the patient? S chorda tympani being burnt. It is unknown if the patient was treated or if the intended procedure was completed.
Patient Sequence No: 1, Text Type: D, B5


[134242960] Olympus received additional information that states the device was held 0. 5mm to 1mm from the patient? S ear during activation. There was no bipolar cautery required for this procedure. The doctor reported that the patient has recovered well.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2951238-2019-00352
MDR Report Key8234422
Report SourceFOREIGN,USER FACILITY
Date Received2019-01-10
Date of Report2019-08-01
Date of Event2018-12-13
Date Mfgr Received2019-07-10
Date Added to Maude2019-01-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. CONNIE TUBERA
Manufacturer Street2400 RINGWOOD AVENUE
Manufacturer CitySAN JOSE CA 95131
Manufacturer CountryUS
Manufacturer Postal95131
Manufacturer Phone4089355124
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSMART STAPES PISTON .6 X 4.50 MM
Generic NameSMART STAPES PISTON
Product CodeETB
Date Received2019-01-10
Model Number70145928
Lot NumberMH594774
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerGYRUS ACMI, INC
Manufacturer Address136 TURNPIKE ROAD SOUTHBOROUGH MA 01772 US 01772


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2019-01-10