WALLFLEX COLONIC M00565040 6504

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2019-01-10 for WALLFLEX COLONIC M00565040 6504 manufactured by Boston Scientific Corporation.

Event Text Entries

[132783517] Patient's exact age is unknown; however it was reported that the patient was over the age of 18. (b)(4). According to the complainant, the suspect device has been disposed and is not available for return. If any further relevant information is received, a supplemental mdr will be filed.
Patient Sequence No: 1, Text Type: N, H10


[132783518] It was reported to boston scientific corporation on (b)(6) 2018 that a wallflex colonic stent was to be used to treat a 4cm malignant stricture in the colon due to colorectal cancer during a lower gastrointestinal stent placement procedure performed on (b)(6) 2018. Reportedly, the patient's anatomy was tortuous and was not dilated prior to stent placement. According to the complainant, during the procedure, the stent prematurely deployed and was free floating in the lumen. The stent was removed with forceps and another wallflex colonic stent was placed to complete the procedure. There were no patient complications as a result of this event.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005099803-2018-62559
MDR Report Key8237608
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2019-01-10
Date of Report2019-01-10
Date of Event2018-12-19
Date Mfgr Received2018-12-19
Device Manufacturer Date2017-07-08
Date Added to Maude2019-01-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactCAROLE MORLEY
Manufacturer Street300 BOSTON SCIENTIFIC WAY
Manufacturer CityMARLBOROUGH MA 01752
Manufacturer CountryUS
Manufacturer Postal01752
Manufacturer Phone5086834015
Manufacturer G1BOSTON SCIENTIFIC IRELAND LIMITED
Manufacturer StreetBALLYBRIT BUSINESS PARK
Manufacturer CityGALWAY
Manufacturer CountryEI
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameWALLFLEX COLONIC
Generic NameSTENT, COLONIC, METALIC, EXPANDABLE
Product CodeMQR
Date Received2019-01-10
Model NumberM00565040
Catalog Number6504
Lot Number0020859767
Device Expiration Date2019-07-08
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBOSTON SCIENTIFIC CORPORATION
Manufacturer Address300 BOSTON SCIENTIFIC WAY MARLBOROUGH MA 01752 US 01752


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2019-01-10