BBL SENSI-DISC TICARCILLIN TIC-75 * 231619

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2007-02-23 for BBL SENSI-DISC TICARCILLIN TIC-75 * 231619 manufactured by Bd Caribe, Ltd..

Event Text Entries

[591007] Customer called to report false susceptible with e. Coli atcc #35218. The zone size should be 6mm and the customer is getting 23mm. Customer is following proper set up procedure and storage.
Patient Sequence No: 1, Text Type: D, B5


[7819018] Internal investigation of retention product also exhibit a false susceptible result when tested against e. Coli atcc 35218. Internal investigation has shown that this lot of ticarcillin contains clavulanic acid, a compound that inactivates e. Coli atcc 35218's beta-lactamase enzymes, allowing the antibiotic to kill the organism creating a zone of growth inhibition with the kirby-bauer test methodology.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1119779-2007-00002
MDR Report Key824841
Report Source05
Date Received2007-02-23
Date of Report2007-02-19
Date of Event2006-11-15
Date Mfgr Received2006-11-15
Device Manufacturer Date2006-06-01
Date Added to Maude2007-03-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer ContactJOHN GERLICH
Manufacturer Street7 LOVETON CIR.
Manufacturer CitySPARKS MD 21152
Manufacturer CountryUS
Manufacturer Postal21152
Manufacturer Phone4103164464
Manufacturer G1*
Manufacturer Street*
Manufacturer City*
Manufacturer Country*
Single Use3
Remedial ActionRC
Previous Use Code3
Removal Correction Number1119779-02/19/07-R001
Event Type3
Type of Report3

Device Details

Brand NameBBL SENSI-DISC TICARCILLIN TIC-75
Generic NameANTIMICROBIAL SUSCEPTIBILITY TEST DISC
Product CodeJTN
Date Received2007-02-23
Returned To Mfg2006-11-22
Model Number*
Catalog Number231619
Lot Number6167435
ID Number*
Device Expiration Date2008-06-30
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key812240
ManufacturerBD CARIBE, LTD.
Manufacturer Address* CAYEY PR * US
Baseline Brand NameBBL SENSI-DISC TICARCILLIN TIC-75
Baseline Generic NameANTIMICROBIAL SUSCEPTIBILITY TEST DISC
Baseline Model No*
Baseline Catalog No231619
Baseline ID*


Patients

Patient NumberTreatmentOutcomeDate
101. Life Threatening 2007-02-23

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.