GLUC3 GLUCOSE HK GEN.3 05168791190

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2019-01-15 for GLUC3 GLUCOSE HK GEN.3 05168791190 manufactured by Roche Diagnostics.

Event Text Entries

[133198113] .
Patient Sequence No: 1, Text Type: N, H10


[133198114] The customer stated that they have sporadically been getting low results for an unspecified number of patient samples tested with gluc3 glucose hk gen. 3 on a cobas 8000 c 702 module. Some samples will have results of 18, 5, or 2. When repeated, these samples have normal results. The customer provided a specific example for one patient sample with an erroneous result. No erroneous results were reported outside of the laboratory. The unit of measure used for gluc3 was not provided. The sample initially resulted with a value of 18, which repeated around 100. No adverse events were alleged to have occurred with the patient. The serial number of the c 702 module is (b)(4). Quality controls were acceptable. The customer declined a service visit. The customer replaced the reagent pack and stated that the issue was resolved. The investigation could not identify a product problem. The cause of the event could not be determined.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2019-00201
MDR Report Key8249418
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2019-01-15
Date of Report2019-01-15
Date of Event2019-01-03
Date Mfgr Received2019-01-03
Date Added to Maude2019-01-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameGLUC3 GLUCOSE HK GEN.3
Generic NameHEXOKINASE, GLUCOSE
Product CodeCFR
Date Received2019-01-15
Model NumberNA
Catalog Number05168791190
Lot Number342842
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2019-01-15