EVOLUTION? COLONIC CONTROLLED-RELEASE STENT - UNCOVERED G48028 EVO-25-30-8-C

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2019-01-16 for EVOLUTION? COLONIC CONTROLLED-RELEASE STENT - UNCOVERED G48028 EVO-25-30-8-C manufactured by Cook Ireland Ltd.

Event Text Entries

[133220464] 510(k) number: k163468. (b)(4). Investigation is still pending, a follow up mdr will be submitted to include the investigation conclusions.
Patient Sequence No: 1, Text Type: N, H10


[133220465] Event is fda mdr reportable based on the device malfunction reporting precedence for this device family for the issue of? Flexor kinked/ stretched/ broken/ compressed'. "impossible to deploy stent". Rep will arrange return. "as per complaint form": stent would not deploy. Additional information received 19-dec-18: was the directional button fully engaged? Yes. Did the red indicator move when the trigger was pressed? Yes. Did the red indicator continue to move after the delivery stopped? Yes. Were any cracking/popping sounds heard from the handle? Not known. Was the stent partially exposed? No. If the stent was partially exposed, was it possible to recapture the stent fully before removal? Na. Were any additional procedures needed? No.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3001845648-2019-00021
MDR Report Key8252017
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2019-01-16
Date of Report2019-02-13
Date of Event2018-12-12
Date Facility Aware2019-01-16
Date Mfgr Received2018-12-12
Device Manufacturer Date2017-01-06
Date Added to Maude2019-01-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. HEATHER RYAN
Manufacturer StreetO HALLORAN ROAD NATIONAL TECHNOLOGY PARK
Manufacturer CityLIMERICK
Manufacturer Phone061334440
Manufacturer G1COOK IRELAND LTD
Manufacturer StreetO HALLORAN ROAD NATIONAL TECHNOLOGY PARK
Manufacturer CityLIMERICK
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameEVOLUTION? COLONIC CONTROLLED-RELEASE STENT - UNCOVERED
Generic NameMQR STENT, COLONIC METALLIC EXPANDABLE
Product CodeMQR
Date Received2019-01-16
Returned To Mfg2019-01-29
Model NumberG48028
Catalog NumberEVO-25-30-8-C
Lot NumberC1311877
Device Expiration Date2019-01-06
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Age705 DA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerCOOK IRELAND LTD
Manufacturer AddressO HALLORAN ROAD LIMERICK


Patients

Patient NumberTreatmentOutcomeDate
10 2019-01-16

© 2026 FDA.report