CARESTREAM DRX REVOLUTION MOBILE X-RAY SYSTEM DRX R-1 1019397

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2019-01-16 for CARESTREAM DRX REVOLUTION MOBILE X-RAY SYSTEM DRX R-1 1019397 manufactured by Carestream Health Inc.

Event Text Entries

[133192510] Per the investigation, there was no device malfunction. This issue was caused by the technician placing the lead marker on the incorrect side of the image. This lead to the chest tube being placed on the incorrect side.
Patient Sequence No: 1, Text Type: N, H10


[133192511] Carestream health received a report related to a chest tube being placed on the incorrect side of the patient. Reporters narrative: an ap (anteroposterior) one view chest was taken and the image came up as pa (posterioranterior). The tech stated that the right lead marker was either collimator out or masked out, unsure. Tech then added a digital left marker on what she thought was the left side. But this was not correct because the image was flipped. The image was sent to the radiologist to be read who did not catch this error and read it incorrectly. The er (emergency room) physician read the image and placed a chest tube on the incorrect side.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1317307-2019-00001
MDR Report Key8252264
Report SourceHEALTH PROFESSIONAL
Date Received2019-01-16
Date of Report2019-01-16
Date of Event2018-12-24
Date Mfgr Received2018-12-31
Date Added to Maude2019-01-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR LEE MILLWARD
Manufacturer Street150 VERONA STREET
Manufacturer CityROCHESTER 14608
Manufacturer CountryUS
Manufacturer Postal14608
Manufacturer G1CARESTREAM HEALTH INC
Manufacturer Street1049 RIDGE ROAD
Manufacturer CityROCHESTER 14615
Manufacturer CountryUS
Manufacturer Postal Code14615
Single Use3
Remedial ActionOT
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCARESTREAM DRX REVOLUTION MOBILE X-RAY SYSTEM
Generic NameDRX REVOLUTION
Product CodeIZL
Date Received2019-01-16
Model NumberDRX R-1
Catalog Number1019397
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerCARESTREAM HEALTH INC
Manufacturer Address150 VERONA STREET ROCHESTER NY 14608 US 14608


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2019-01-16