I-STAT CG4+ CARTRIDGE 03P85-50

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2019-01-16 for I-STAT CG4+ CARTRIDGE 03P85-50 manufactured by Abbott Point Of Care.

Event Text Entries

[134579781] Apoc incident # (b)(4). Apoc labeling will be evaluated during the investigation as pertaining to the event.
Patient Sequence No: 1, Text Type: N, H10


[134579782] On (b)(6) 2018, abbott point of care was contacted by a customer regarding i-stat cg4+ cartridges (lot# unk) that yielded suspected discrepant pco2 results during a study of hypothermic babies. There was no patient information available at the time of this report. The customer stated that they are not live for testing at this time and that the results were not used to treat the patient. (b)(6). Test and collection times not available. There are no injuries associated with this event. At this time there is no reason to suspect a malfunction exists. The reporting decision was based on the limited information available that suggests the product was not performing within the variability of the assay. The investigation is underway.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2245578-2019-00011
MDR Report Key8253294
Report SourceHEALTH PROFESSIONAL
Date Received2019-01-16
Date of Report2019-04-01
Date of Event2018-12-26
Date Mfgr Received2019-03-26
Date Added to Maude2019-01-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactLINDA MACZUSZENKO
Manufacturer Street400 COLLEGE ROAD
Manufacturer CityPRINCETON NJ 08540
Manufacturer CountryUS
Manufacturer Postal08540
Manufacturer Phone6136885949
Manufacturer G1ABBOTT POINT OF CARE CANADA LTD.
Manufacturer Street185 CORKSTOWN ROAD
Manufacturer CityOTTAWA, ONTARIO K2H 8V4
Manufacturer CountryCA
Manufacturer Postal CodeK2H 8V4
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameI-STAT CG4+ CARTRIDGE
Generic NameCG4+ CARTRIDGE
Product CodeKHP
Date Received2019-01-16
Model NumberNA
Catalog Number03P85-50
Lot NumberUNK
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerABBOTT POINT OF CARE
Manufacturer Address400 COLLEGE ROAD PRINCETON NJ 085406607 US 085406607


Patients

Patient NumberTreatmentOutcomeDate
10 2019-01-16