SYSTEM 1E PROCESSOR

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-01-16 for SYSTEM 1E PROCESSOR manufactured by Steris Canada Corporation.

Event Text Entries

[133326890] A steris service technician arrived onsite to inspect the processor and found the filter housing assembly for the water filtration system was cracked allowing water to leak out onto the floor. The user facility replaced the filter housing. The technician tested the unit, confirmed it to be operating according to specification, and returned it to service. No additional issues have been reported.
Patient Sequence No: 1, Text Type: N, H10


[133326891] The user facility reported that their system 1e processor was leaking water onto the floor. No report of injury.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9680353-2019-00003
MDR Report Key8254717
Date Received2019-01-16
Date of Report2019-01-16
Date of Event2018-12-27
Date Mfgr Received2018-12-27
Date Added to Maude2019-01-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. DANIEL DAVY
Manufacturer Street5960 HEISLEY ROAD
Manufacturer CityMENTOR OH 44060
Manufacturer CountryUS
Manufacturer Postal44060
Manufacturer Phone4403927453
Manufacturer G1STERIS CANADA CORPORATION
Manufacturer Street490, ARMAND-PARIS
Manufacturer CityQUEBEC, QUEBEC GIC 8A3
Manufacturer CountryCA
Manufacturer Postal CodeGIC 8A3
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameSYSTEM 1E PROCESSOR
Generic NameSYSTEM 1E PROCESSOR
Product CodeMED
Date Received2019-01-16
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSTERIS CANADA CORPORATION
Manufacturer Address490 ARMAND-PARIS QUEBEC, QUEBEC GIC 8A3 CA GIC 8A3


Patients

Patient NumberTreatmentOutcomeDate
10 2019-01-16

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.