LACTATE DEHYDROGENASE 02P56-21

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional report with the FDA on 2019-01-18 for LACTATE DEHYDROGENASE 02P56-21 manufactured by Abbott Manufacturing Inc.

Event Text Entries

[134746477] This issue was previously reported under mdr number 1415939-2018-00152. The incorrect manufacturing location was documented in the original report. This report was generated to document (b)(4) as the manufacture location. (b)(6). Further investigation of the customer issue included a review of the complaint text, an instrument log review, a search for similar complaints, design history record (dhr) review, and a review of labeling. Return material was not available. A review of tracking and trending did not identify an adverse trend for issue observed by the customer. The dhr review did not identify any potential non-conformances, non-conformances or deviations related to the complaint issue. Labeling was reviewed and found to be adequate. Based on the available information no deficiency of the clinical chemistry ldh assay, reagent list number 02p56, lot 99549un18, was identified.
Patient Sequence No: 1, Text Type: N, H10


[134746478] The customer observed falsely elevated lactate dehydrogenase (ldh) results while using the clinical chemistry ldh assay. The customer provided the following results for 2 patients (reference range 125 to 220 u/l): patient 1: initial 303, retests: 238, 218, 300 and 295 u/l; patient 2: initial 349, retests: 214, 205 and 206. No impact to patient management was reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1628664-2019-00040
MDR Report Key8261501
Report SourceFOREIGN,HEALTH PROFESSIONAL
Date Received2019-01-18
Date of Report2019-01-18
Date Mfgr Received2019-01-08
Device Manufacturer Date2018-05-01
Date Added to Maude2019-01-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNOEMI ROMERO-KONDOS, RN BSN
Manufacturer Street100 ABBOTT PARK ROAD DEPT. 09B9, LCCP1-3
Manufacturer CityABBOTT PARK IL 600643537
Manufacturer CountryUS
Manufacturer Postal600643537
Manufacturer Phone224667-512
Manufacturer G1ABBOTT MANUFACTURING INC
Manufacturer Street1921 HURD DRIVE
Manufacturer CityIRVING TX 75038
Manufacturer CountryUS
Manufacturer Postal Code75038
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameLACTATE DEHYDROGENASE
Generic NameLACTATE DEHYDROGENASE
Product CodeCFJ
Date Received2019-01-18
Catalog Number02P56-21
Lot Number99549UN18
Device Expiration Date2019-02-19
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerABBOTT MANUFACTURING INC
Manufacturer Address1921 HURD DRIVE IRVING TX 75038 US 75038


Patients

Patient NumberTreatmentOutcomeDate
10 2019-01-18

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.