LOVE-GRUENWALD PITUITARY FORCEPS 2X10MM NL6146

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-01-18 for LOVE-GRUENWALD PITUITARY FORCEPS 2X10MM NL6146 manufactured by Carefusion, Inc.

Event Text Entries

[135508655] (b)(4) on (b)(6) 2019, writer sent the customer an email acknowledging receipt of the complaint, providing the complaint tracking number and requested follow up information including as to if there was any patient impact related to this event. Writer provided contact information. Device not yet evaluated, if the device is evaluated a follow up will be sent.
Patient Sequence No: 1, Text Type: N, H10


[135508656] It was reported via email: yesterday one of our pituitary? S had a piece of metal come off during a case. The tech saw it and it wasn? T left in the patient, but we had to take it out of service. What was the procedure that was being performed? Cervical disc replacement. Did any part the instrument fall into the patient? S body, and if so, how was it retrieved? Unknown. Was there a medical procedure performed to verify if the instrument was in the patient? S body, such as an x-ray? X-ray down, nothing seen on x-ray. What was the patient? S outcome? Stable, normal outcome. Was the procedure completed as planned? Yes. Can you please send all parts of the instrument for evaluation? Yes, metal piece was found and matched to instrument. Do you have the lot number? J17xsx. Could you please clarify where was the broken piece found? It was found on a sponge by the scrub tech after wiping the instrument after normal use. No further information available.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1423507-2019-00001
MDR Report Key8262616
Date Received2019-01-18
Date of Report2019-03-01
Date of Event2019-01-09
Date Mfgr Received2019-01-10
Date Added to Maude2019-01-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactBDX ANNA WEHRHEIM
Manufacturer Street75 NORTH FAIRWAY DRIVE
Manufacturer CityVERNON HILLS IL 60061
Manufacturer CountryUS
Manufacturer Postal60061
Manufacturer G1CAREFUSION, INC
Manufacturer Street75 NORTH FAIRWAY DRIVE
Manufacturer CityVERNON HILLS IL 60061
Manufacturer CountryUS
Manufacturer Postal Code60061
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameLOVE-GRUENWALD PITUITARY FORCEPS 2X10MM
Generic NameKERRISON RONGEURS
Product CodeHAE
Date Received2019-01-18
Catalog NumberNL6146
Lot NumberJ17XSX
OperatorHEALTH PROFESSIONAL
Device Availability*
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerCAREFUSION, INC
Manufacturer Address75 NORTH FAIRWAY DRIVE VERNON HILLS IL 60061 US 60061


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2019-01-18

© 2026 FDA.report