APOLLO WHIRLPOOL BATH 6000

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-01-21 for APOLLO WHIRLPOOL BATH 6000 manufactured by Apollo Corporation.

Event Text Entries

[133846499] Through inspecting the device it was determined that the incident was caused by the carrier base being out of adjustment. The pin retainer dropped down below its recommended height, preventing the unit from latching properly. It was noted that the docking pin retainer, one load cell, and 3 of the casters were all loose on the unit. The loosening of the pin retainer is likely what led to it falling out of adjustment. All components were tightened and brought back into adjustment by an apollo representative. The representative discussed planned maintenance requirements with the facility's maintenance and administration personnel to prevent reoccurrence of this event.
Patient Sequence No: 1, Text Type: N, H10


[133846500] Bath aide was preparing to give resident a bath. As the resident was being transferred into tub, the scale base separated from the bath tub and the resident and chair fell between the scale base and the tub. The resident hit their head on side of tub during the fall, causing a contusion and 3 small abrasions on head.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2182947-2019-00001
MDR Report Key8264892
Date Received2019-01-21
Date of Report2019-01-21
Date of Event2018-12-21
Device Manufacturer Date2014-05-19
Date Added to Maude2019-01-21
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. RANDALL DEKAN
Manufacturer Street450 MAIN STREET
Manufacturer CitySOMERSET WI 54025
Manufacturer CountryUS
Manufacturer Postal54025
Manufacturer Phone7152475625
Single Use3
Remedial ActionMA
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameAPOLLO WHIRLPOOL BATH
Generic NameBATH CHAIR
Product CodeILJ
Date Received2019-01-21
Model Number6000
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerAPOLLO CORPORATION
Manufacturer Address450 MAIN STREET SOMERSET WI 54025 US 54025


Patients

Patient NumberTreatmentOutcomeDate
10 2019-01-21

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.