MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2019-01-22 for COOK CERVICAL RIPENING BALLOON W/STYLET J-CRBS-184000 manufactured by Cook Inc.
[134185909]
Occupation: other healthcare professional- assistant manager (b)(4). This report includes information known at this time. A follow up report will be submitted should additional relevant information become available.
Patient Sequence No: 1, Text Type: N, H10
[134185910]
It was reported that during a "foley bulb induction of cervix" using a cook cervical ripening balloon w/stylet; when inserting the ripening balloon, the certified nurse midwife pushed on the stylet, felt it bend, as well as felt resistance in the stylet. When it was pulled out, the stylet was found to have punctured the balloon at the blue plastic tip of the device. The entire balloon was removed, and another unspecified device was used to complete the procedure successfully. No adverse events have been reported as a result of the alleged malfunction.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 1820334-2019-00043 |
MDR Report Key | 8266547 |
Report Source | HEALTH PROFESSIONAL,USER FACI |
Date Received | 2019-01-22 |
Date of Report | 2019-01-30 |
Date of Event | 2019-01-04 |
Date Mfgr Received | 2019-01-26 |
Device Manufacturer Date | 2018-10-11 |
Date Added to Maude | 2019-01-22 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Reporter Occupation | OTHER HEALTH CARE PROFESSIONAL |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MR. LARRY POOL |
Manufacturer Street | 750 DANIELS WAY |
Manufacturer City | BLOOMINGTON IN 47404 |
Manufacturer Country | US |
Manufacturer Postal | 47404 |
Manufacturer Phone | 8123392235 |
Manufacturer G1 | COOK INC |
Manufacturer Street | 750 DANIELS WAY |
Manufacturer City | BLOOMINGTON IN 47404 |
Manufacturer Country | US |
Manufacturer Postal Code | 47404 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | COOK CERVICAL RIPENING BALLOON W/STYLET |
Generic Name | PFJ, DILATOR, CERVICAL, HYGROSCOPIC-IAMINARIA |
Product Code | PFJ |
Date Received | 2019-01-22 |
Returned To Mfg | 2019-01-22 |
Catalog Number | J-CRBS-184000 |
Lot Number | 9227866 |
Operator | HEALTH PROFESSIONAL |
Device Availability | R |
Device Age | DA |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | COOK INC |
Manufacturer Address | 750 DANIELS WAY BLOOMINGTON IN 47404 US 47404 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2019-01-22 |