DIGITAL RADIOGRAPHY X-RAY SYSTEM DX-D100

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-01-22 for DIGITAL RADIOGRAPHY X-RAY SYSTEM DX-D100 manufactured by Agfa N.v..

Event Text Entries

[133954714] Agfa healthcare (b)(4) has recently changed the company structure into two legal entities, named agfa healthcare and agfa (b)(4). This report is for agfa (b)(4). (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[133954715] On january 7, 2019, a customer in (b)(6) reported to agfa that when using their dx-d 100 system, they experienced unintended movement. The customer reported the dx-d 100 system was moved into position to take an image when the unit made a long beeping sound and then started to move forward with no one touching the handle. The dx-d 100 hit a wall but was still trying to go forward. The system stopped when the emergency stop button was pressed. Investigation is under way and a supplemental report will be provided. There has been no reported harm to patient or user during these events.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9616389-2019-00005
MDR Report Key8266697
Date Received2019-01-22
Date of Report2019-03-04
Date of Event2019-01-07
Date Mfgr Received2019-02-19
Date Added to Maude2019-01-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS. CASSANDRA MCGOWAN
Manufacturer Street10 S ACADEMY STREET
Manufacturer CityGREENVILLE SC 29601
Manufacturer CountryUS
Manufacturer Postal29601
Manufacturer Phone8644211984
Manufacturer G1AGFA N.V.
Manufacturer StreetSEPTESTRAAT 27
Manufacturer CityMORTSEL, B2640
Manufacturer CountryBE
Manufacturer Postal CodeB2640
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameDIGITAL RADIOGRAPHY X-RAY SYSTEM DX-D100
Generic NameDX-D 100 MOBILE
Product CodeIZL
Date Received2019-01-22
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerAGFA N.V.
Manufacturer AddressSEPTESTRAAT 27 MORTSEL, B2640 BE B2640


Patients

Patient NumberTreatmentOutcomeDate
10 2019-01-22

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.