Event Text Entries

[134595155] The hill-rom technical support representative performed troubleshooting over the phone with the account, asking the account to check the side comm board. It is necessary for the compella bariatric bed system to have an effective maintenance program. We recommend that you do annual preventive maintenance. Specified checks: the communication cable and its connectors are in good condition. The sidecom communication system operates correctly. If any of these checks fail, repair or replace the part as applicable. A search of the hill-rom maintenance records did not show hill-rom performed any preventative maintenance on this bed. It is unknown if the facility performs preventative maintenance on their beds. The account replaced the side comm board to resolve the issue. Based on this information, no further action is required.
Patient Sequence No: 1, Text Type: N, H10


[134595156] Hill-rom received a report from the account stating the nurse call was not working. The bed was located at the account. There was no patient/user injury reported. This report was filed in our complaint handling system as complaint (b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1824206-2019-00030
MDR Report Key8271088
Date Received2019-01-23
Date of Report2019-01-03
Date of Event2019-01-03
Date Mfgr Received2019-01-03
Date Added to Maude2019-01-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactEMILY MITCHELL
Manufacturer Street1069 STATE ROUTE 46 EAST
Manufacturer CityBATESVILLE IN 47006
Manufacturer CountryUS
Manufacturer Postal47006
Manufacturer Phone8129310130
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameCOMPELLA CAP US SCALE PD PPM
Generic NameBARIATRIC BED
Product CodeOSI
Date Received2019-01-23
Model NumberP7800A000001
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerHILL-ROM BATESVILLE
Manufacturer Address1069 STATE ROUTE 46 EAST BATESVILLE IN 47006 US 47006


Patients

Patient NumberTreatmentOutcomeDate
10 2019-01-23