SYMMETRY DIETHRICH 32-844

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-01-24 for SYMMETRY DIETHRICH 32-844 manufactured by Symmetry Surgical Inc.

Event Text Entries

[136010587] The device in question was purchased (b)(6) 2017 and had approximately 1. 5 years of use. Upon receiving the returned device, the hardness and material of the device were evaluated. It was determined that the device was within specifications. Upon further evaluation of the device, it was noticed that the cutting edges were scratched and chipped in some place. This type of damage is consistent with damage that would occur if the scissor was used in a way not intended for the purpose of the device. This particular device has an extremely delicate scissor tip that can easily become broken if used incorrectly. This is the only complaint recorded for this device with (b)(4) sold. This report can be seen as the final report. If additional information is received that alleges any additional patient involvement or a need for corrective actions a follow-up report will be submitted.
Patient Sequence No: 1, Text Type: N, H10


[136010588] The tip of the diethrich scissor broke off. It was noticed that the tip was broken in reprocessing. As a result the patient had to be recalled and an investigation was carried out to determine if the piece remained in the patient. It was communicated that the patient was not impacted by the broken tip.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3007208013-2019-00003
MDR Report Key8274807
Date Received2019-01-24
Date of Report2019-01-24
Date of Event2018-11-12
Date Mfgr Received2019-01-07
Date Added to Maude2019-01-24
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactBRANDI MEATH
Manufacturer Street3034 OWEN DRIVE
Manufacturer CityANTIOCH TN 37013
Manufacturer CountryUS
Manufacturer Postal37013
Manufacturer G1SYMMETRY SURGICAL INC
Manufacturer Street3034 OWEN DRIVE
Manufacturer CityANTIOCH TN 37013
Manufacturer CountryUS
Manufacturer Postal Code37013
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameSYMMETRY DIETHRICH
Generic NameSCISSORS
Product CodeLRW
Date Received2019-01-24
Returned To Mfg2019-01-23
Catalog Number32-844
Lot Number110216
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSYMMETRY SURGICAL INC
Manufacturer Address3034 OWEN DRIVE ANTIOCH TN 37013 US 37013


Patients

Patient NumberTreatmentOutcomeDate
10 2019-01-24

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.