BALLOON MAJ-1351

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2019-01-24 for BALLOON MAJ-1351 manufactured by Olympus Medical Systems Corp..

Event Text Entries

[134065123] The subject device was not returned to olympus medical systems corp. (omsc) for evaluation. Therefore, the exact cause of the reported event could not be conclusively determined. The manufacturing record was reviewed and found no irregularities. This type of the event is most likely related to the operator's technique. Based on the past similar cases, it was known that the event occurred due to the following causes. The user could not attach the balloon to ultrasonic bronchoscope correctly. The instruction manual of bronchoscope describes a method "attaching the balloon".
Patient Sequence No: 1, Text Type: N, H10


[134065124] During an unspecified procedure, the subject device was used. In the procedure, when the ultrasonic bronchoscope was pulled out from the patient's body, the subject device was missing from the tip of the bronchoscope. The intended procedure was completed with another device. Doctor did not know whether the device was fell into the patient body or outside the patient body. There was no patient injury reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number8010047-2019-00949
MDR Report Key8276606
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2019-01-24
Date of Report2019-03-07
Date of Event2018-12-27
Date Mfgr Received2019-02-20
Date Added to Maude2019-01-24
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR KAZUTAKA MATSUMOTO
Manufacturer Street2951 ISHIKAWA-CHO
Manufacturer CityHACHIOJI-SHI, TOKYO-TO 192-8507
Manufacturer CountryJA
Manufacturer Postal192-8507
Manufacturer Phone426425177
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameBALLOON
Generic NameBALLOON
Product CodePSV
Date Received2019-01-24
Model NumberMAJ-1351
Lot Number89K
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerOLYMPUS MEDICAL SYSTEMS CORP.
Manufacturer Address2951 ISHIKAWA-CHO HACHIOJI-SHI, TOKYO-TO US


Patients

Patient NumberTreatmentOutcomeDate
10 2019-01-24

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.