EVIS EXERA II ULTRASONIC BRONCHOFIBERVIDEOSCOPE BF-UC180F

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-01-24 for EVIS EXERA II ULTRASONIC BRONCHOFIBERVIDEOSCOPE BF-UC180F manufactured by Olympus Medical Systems Corp..

Event Text Entries

[134982076] The subject device has not been returned to omsc but was returned to olympus (b)(4). (b)(4) sent the device to a third party laboratory for microbiological testing. As a result of the testing, no microbe was detected from the sample collected from the instrument channel, the air/water channel and the distal end of the device. Omsc reviewed the manufacturing history of the subject device and confirmed no irregularity. The exact cause of the reported event could not be conclusively determined at this time. If additional information becomes available, this report will be supplemented.
Patient Sequence No: 1, Text Type: N, H10


[134982077] Olympus medical systems corp. (omsc) was informed that as a result of routine microbiological testing by the user facility, the sample collected from instrument channel of the subject device tested positive for staphylococcus sp. The user facility did not provide other detailed information such as the number of microbes. The device had been reprocessed with a non-olympus automated endoscope reprocessor, belimed. There was no report of infection associated with this report.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number8010047-2019-00957
MDR Report Key8276719
Date Received2019-01-24
Date of Report2019-01-25
Date Mfgr Received2019-01-09
Device Manufacturer Date2016-04-07
Date Added to Maude2019-01-24
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR KAZUTAKA MATSUMOTO
Manufacturer Street2951 ISHIKAWA-CHO
Manufacturer CityHACHIOJI-SHI, TOKYO-TO 192-8507
Manufacturer CountryJA
Manufacturer Postal192-8507
Manufacturer Phone426425177
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameEVIS EXERA II ULTRASONIC BRONCHOFIBERVIDEOSCOPE
Generic NameULTRASONIC BRONCHOFIBERVIDEOSCOPE
Product CodePSV
Date Received2019-01-24
Model NumberBF-UC180F
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerOLYMPUS MEDICAL SYSTEMS CORP.
Manufacturer Address2951 ISHIKAWA-CHO HACHIOJI-SHI, TOKYO-TO 192-8507 JA 192-8507


Patients

Patient NumberTreatmentOutcomeDate
10 2019-01-24

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.