STONE CONE M0063903200 390-320

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2019-01-25 for STONE CONE M0063903200 390-320 manufactured by Boston Scientific Corporation.

Event Text Entries

[134084114] (b)(4). The device has not been received for analysis. Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
Patient Sequence No: 1, Text Type: N, H10


[134084115] It was reported to boston scientific corporation that a stone cone retrieval coil was used during a retrograde intrarenal surgery (rirs) procedure performed in the ureter on (b)(6) 2019. The indication for the procedure was for ureterscopic lithotripsy. According to the complainant, during procedure, when lithotripsy was finished the complainant observed that the length of the stone cone seemed shorter than usual. Another stone cone retrieval coil was used for length comparison and it was discovered that the tip of the stone cone that was used had broken off and was in the ureter of the patient. Tools were used to take out the tip from the patient. There was no serious injury, nor were there any adverse patient effects reported as a result of this event.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005099803-2019-00291
MDR Report Key8277406
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2019-01-25
Date of Report2019-03-18
Date of Event2019-01-04
Date Mfgr Received2019-02-25
Device Manufacturer Date2018-06-25
Date Added to Maude2019-01-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactCAROLE MORLEY
Manufacturer Street300 BOSTON SCIENTIFIC WAY ,
Manufacturer CityMARLBOROUGH MA 01752
Manufacturer CountryUS
Manufacturer Postal01752
Manufacturer Phone5086834015
Manufacturer G1LAKE REGION MEDICAL
Manufacturer Street1525 HERTZ STREET
Manufacturer CityCHIHUAHUA
Manufacturer CountryMX
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSTONE CONE
Generic NameDISLODGER, STONE, FLEXIBLE
Product CodeFGO
Date Received2019-01-25
Returned To Mfg2019-02-01
Model NumberM0063903200
Catalog Number390-320
Lot Number0004346450
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBOSTON SCIENTIFIC CORPORATION
Manufacturer Address300 BOSTON SCIENTIFIC WAY , MARLBOROUGH MA 01752 US 01752


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2019-01-25