BIL-D BILIRUBIN DIRECT BIL-D G2

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-01-28 for BIL-D BILIRUBIN DIRECT BIL-D G2 manufactured by Roche Diagnostics.

Event Text Entries

[134334503] The investigation could not identify a product problem. The cause of the event could not be determined. The follow up/corrective actions were that the customer changed their method from the "jendrassic" "groff" method to the doumas method. The doumas method will provide the customer with the expected ranges that they prefer. No devices were returned. This device is not labeled for single use and is not reprocessed or reused.
Patient Sequence No: 1, Text Type: N, H10


[134334504] This report summarizes 1 malfunction event. Erroneous high results were generated by the cobas c 111. The events involved a total of 44 patients with high results for the bil-d bilirubin direct assay. The patients' ages ranged from (b)(6) - (b)(6) years. The other patients' ages were requested, but were not provided. The patients' weights were requested, but were not provided. There were 3 males and 6 females. The other patients' genders were requested, but were not provided. The patients' races were requested, but were not provided. The patients' ethnicities were requested, but were not provided.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2019-90008
MDR Report Key8280716
Date Received2019-01-28
Date Added to Maude2019-01-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA0
Report to FDA3
Event Location3
Manufacturer ContactMICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG), NA 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameBIL-D BILIRUBIN DIRECT
Generic NameDIAZO COLORIMETRY, BILIRUBIN
Product CodeCIG
Date Received2019-01-28
Model NumberBIL-D G2
Lot NumberASKU
Device Availability*
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 46250 US 46250


Patients

Patient NumberTreatmentOutcomeDate
10 2019-01-28