TOTAL PROTEIN URINE/CSF GEN.3 TPUC G3

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-01-28 for TOTAL PROTEIN URINE/CSF GEN.3 TPUC G3 manufactured by Roche Diagnostics.

Event Text Entries

[134461867] The investigation did not identify a product problem. The cause of the event could not be determined. The follow up/corrective actions events were that the customer loaded a new pack of reagent, calibrated the system, and performed successful qc testing. Following the loading of a new reagent pack, the customer was satisfied with the results from the system. No devices were returned. This device is not labeled for single use and is not reprocessed or reused.
Patient Sequence No: 1, Text Type: N, H10


[134461868] This report summarizes 1 malfunction event. Erroneous high results were generated by the cobas 8000 c 502 module compared to both the electrophoresis results and another cobas c502. The events involved a total of 12 patients with erroneous results for tpuc3 total protein urine/csf gen. 3. The patients' ages were requested but were not provided. The patients' weights were requested but were not provided. The patients' genders were requested but were not provided. The patients' races were requested but were not provided the patients' ethnicities were requested but were not provided.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2019-90002
MDR Report Key8281607
Date Received2019-01-28
Date Added to Maude2019-01-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA0
Report to FDA3
Event Location3
Manufacturer ContactMICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG), NA 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameTOTAL PROTEIN URINE/CSF GEN.3
Generic NameTURBIDIMETRIC, TOTAL PROTEIN
Product CodeJGQ
Date Received2019-01-28
Model NumberTPUC G3
Lot NumberASKU
Device Availability*
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 46250 US 46250


Patients

Patient NumberTreatmentOutcomeDate
10 2019-01-28

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.