LDHI2 LACTATE DEHYDROGENASE ACC. TO IFCC VER.2 LDHI G2

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-01-28 for LDHI2 LACTATE DEHYDROGENASE ACC. TO IFCC VER.2 LDHI G2 manufactured by Roche Diagnostics.

Event Text Entries

[134469708] The investigation did not identify a product problem. The cause of the event could not be determined. The follow up/corrective actions were that the customer changed the reagent cassette and the issue seems to be resolved as there have been no further issues. Precision check testing was performed with acceptable results. No devices were returned this device is not labeled for single use and is not reprocessed or reused.
Patient Sequence No: 1, Text Type: N, H10


[134469709] This report summarizes 1 malfunction event. An erroneous low result was generated by the cobas 8000 c 702 module. The event involved 1 patient with an erroneous result for ldhi2 lactate dehydrogenase acc. To ifcc ver. 2. The patient's age was requested but was not provided. The patient's weight was requested but was not provided. The patient's gender was requested but was not provided. The patient's race was requested but was not provided. The patient's ethnicity was requested but was not provided.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2019-90010
MDR Report Key8281628
Date Received2019-01-28
Date Added to Maude2019-01-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA0
Report to FDA3
Event Location3
Manufacturer ContactMICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG), NA 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameLDHI2 LACTATE DEHYDROGENASE ACC. TO IFCC VER.2
Generic NameLACTATE DEHYDROGENASE TEST SYSTEM
Product CodeCFJ
Date Received2019-01-28
Model NumberLDHI G2
Lot Number36247301
Device Availability*
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 46250 US 46250


Patients

Patient NumberTreatmentOutcomeDate
10 2019-01-28