COVERED MOUNTED CP STENT 428 CMCP025

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-01-28 for COVERED MOUNTED CP STENT 428 CMCP025 manufactured by Numed, Inc..

Event Text Entries

[135522669] The complaint stent, bib and ring mandril were returned in separate (b)(6) bags. All of the returned components are clean and free of blood. The covering is detached and was not returned. The adhesive spots are visible on the stent. The stent has been slightly flattened. The catheter was returned with (2) 0. 015 mandrils in the guidewire lumen. There is no blood on the catheter and no contrast in the balloon. The balloon has very well defined folds. The condition of the stent and the bib leads us to believe that they never entered he introducer. The stent and catheter are both completely free of blood and the balloon still has well defined folds that would not be present if the "poba" had occurred. At this time it is impossible to determine the actual nature of the complaint, which makes it impossible to determine the root cause. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[135522670] As reported to numed - "in the first instance we were able to do 'poba' only as the stent remained in the sheath itself. "
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1318694-2019-00004
MDR Report Key8282446
Date Received2019-01-28
Date of Report2019-01-28
Date of Event2018-12-06
Date Mfgr Received2019-01-03
Device Manufacturer Date2017-10-17
Date Added to Maude2019-01-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. NICHELLE LAFLESH
Manufacturer Street2880 MAIN STREET
Manufacturer CityHOPKINTON NY 12965
Manufacturer CountryUS
Manufacturer Postal12965
Manufacturer Phone3153284491
Manufacturer G1NUMED, INC.
Manufacturer Street2880 MAIN STREET
Manufacturer CityHOPKINTON NY 12965
Manufacturer CountryUS
Manufacturer Postal Code12965
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameCOVERED MOUNTED CP STENT
Generic NameAORTIC STENT
Product CodePNF
Date Received2019-01-28
Returned To Mfg2019-01-10
Model Number428
Catalog NumberCMCP025
Lot NumberCMCP-1756
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerNUMED, INC.
Manufacturer Address2880 MAIN STREET HOPKINTON NY 12965 US 12965


Patients

Patient NumberTreatmentOutcomeDate
10 2019-01-28

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.