TUR IRRIGATION SET 2C4041

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-01-28 for TUR IRRIGATION SET 2C4041 manufactured by Baxter Healthcare - Cartago.

Event Text Entries

[134366961] (b)(4). The device was received for evaluation. The returned unit was labeled for single use. Visual inspection revealed that both clamps were missing from the set. The reported condition was verified. The cause of the condition was determined to be an assembly issue during manufacturing. A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot. Should additional relevant information become available, a supplemental report will be submitted.
Patient Sequence No: 1, Text Type: N, H10


[134366962] This report summarizes 1 malfunction events. It was reported that both clamps were missing from a y-type tur/bladder irrigation set. This was identified during priming. There was no patient involvement. No additional information is available.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1416980-2019-00322
MDR Report Key8283050
Date Received2019-01-28
Date of Report2019-01-28
Date Mfgr Received2018-12-31
Date Added to Maude2019-01-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA0
Report to FDA3
Event Location3
Manufacturer Street25212 W. ILLINOIS ROUTE 120
Manufacturer CityROUND LAKE IL 60073
Manufacturer CountryUS
Manufacturer Postal60073
Manufacturer Phone2242702068
Manufacturer G1BAXTER HEALTHCARE - CARTAGO
Manufacturer StreetSEE H10
Manufacturer CityCARTAGO 30106
Manufacturer CountryCS
Manufacturer Postal Code30106
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameTUR IRRIGATION SET
Generic NameSYSTEM, IRRIGATION, UROLOGICAL
Product CodeLJH
Date Received2019-01-28
Model NumberNA
Catalog Number2C4041
Lot NumberR17K29024
Device Availability*
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerBAXTER HEALTHCARE - CARTAGO
Manufacturer AddressCARTAGO


Patients

Patient NumberTreatmentOutcomeDate
10 2019-01-28

© 2026 FDA.report