[134462649] The single-use device was received for evaluation. Visual inspection did not identify any abnormalities that could have contributed to the reported condition. Leak testing and clear passage testing were performed with no issues noted. The device was found to flow normally with no defects or leaks observed. The reported condition was not verified. A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot. Should additional relevant information become available, a supplemental report will be submitted.
Patient Sequence No: 1, Text Type: N, H10


[134462650] This report summarizes 1 malfunction events. It was reported that a micro-volume extension set leaked during infusion. The event involved a patient with no patient consequences. No additional information is available.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1416980-2019-00366
MDR Report Key8287444
Date Received2019-01-29
Date of Report2019-01-29
Date Mfgr Received2018-12-31
Date Added to Maude2019-01-29
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA0
Report to FDA3
Event Location3
Manufacturer Street25212 W. ILLINOIS ROUTE 120
Manufacturer CityROUND LAKE IL 60073
Manufacturer CountryUS
Manufacturer Postal60073
Manufacturer Phone2242702068
Manufacturer G1BAXTER HEALTHCARE - DOMINICAN REPUBLIC
Manufacturer StreetCARRETERA SANCHEZ KM 18.5 PARQUE INDUSTRIAL ITABO, PIISA
Manufacturer CityHAINA, SAN CRISTOBAL 91000
Manufacturer CountryDR
Manufacturer Postal Code91000
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameMICRO-VOLUME EXTENSION SET
Generic NameFILTER, INFUSION LINE
Product CodeFPB
Date Received2019-01-29
Model NumberNA
Catalog Number2N3342
Lot NumberDR18I06024
Device Availability*
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerBAXTER HEALTHCARE - HAINA
Manufacturer AddressHAINA SAN CRISTOBAL


Patients

Patient NumberTreatmentOutcomeDate
10 2019-01-29