EVOLUTION? DUODENAL CONTROLLED-RELEASE STENT - UNCOVERED G48026 EVO-22-27-9-D

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2019-01-31 for EVOLUTION? DUODENAL CONTROLLED-RELEASE STENT - UNCOVERED G48026 EVO-22-27-9-D manufactured by Cook Ireland Ltd.

Event Text Entries

[134778698] 510(k) number: k163468. (b)(4). Investigation is still pending, follow up mdr will be submitted to include the investigation conclusions.
Patient Sequence No: 1, Text Type: N, H10


[134778699] Event is fda mdr reportable based on the device malfunction reporting precedence for this device family for the issue of? Flexor kinked/stretched/broken/compressed?. As reported to customer relations, "we removed the red tab and began to deploy the stent in standard fashion. While deploying it was noted, no evidence of sheath sliding back. Distal aspect of stent was not opening. We arrived at the point of not return and had no evidence of distal stent opening. At this point they carefully examined the device and noticed the blue over sheath was completely sheered where it enters the hub. Unable to deploy the stent and the case was aborted. "
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3001845648-2019-00033
MDR Report Key8292544
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2019-01-31
Date of Report2019-01-31
Date Facility Aware2019-01-08
Date Mfgr Received2019-01-09
Device Manufacturer Date2018-09-11
Date Added to Maude2019-01-31
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. HEATHER RYAN
Manufacturer StreetO HALLORAN ROAD NATIONAL TECHNOLOGY PARK
Manufacturer CityLIMERICK
Manufacturer Phone061334440
Manufacturer G1COOK IRELAND LTD
Manufacturer StreetO HALLORAN ROAD NATIONAL TECHNOLOGY PARK
Manufacturer CityLIMERICK
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameEVOLUTION? DUODENAL CONTROLLED-RELEASE STENT - UNCOVERED
Generic NameMUM STENT, METALIC EXPANDABLE, DUODENAL
Product CodeMUM
Date Received2019-01-31
Returned To Mfg2019-02-05
Model NumberG48026
Catalog NumberEVO-22-27-9-D
Lot NumberC1539098
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerCOOK IRELAND LTD
Manufacturer AddressO HALLORAN ROAD LIMERICK


Patients

Patient NumberTreatmentOutcomeDate
10 2019-01-31

© 2026 FDA.report