[15006341] According to the fire dept. Investigators, an electrical malfunction in a dental x-ray processing machine caused the wiring within the machine to overheat. This overheated wiring ignited part of the machine and nearby combustibles. The x-ray processing machine was a dent x 810 processor machine, installed by mohawk dental supply co. On 10/24/05. The room the equipment was located in was consumed by extensive fire damage. Smoke from the fire extended to other areas in the building and was difficult to ventilate, nearly required evacuation of the attached nursing facility.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report NumberMW1042254
MDR Report Key829338
Date Received2007-03-16
Date of Report2007-03-09
Date of Event2007-03-02
Date Added to Maude2007-03-21
Event Key0
Report Source CodeVoluntary report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameDENT X
Generic NameFILM PROCESSOR
Product CodeEGY
Date Received2007-03-16
Model Number810 PROCESSOR MACH
Catalog Number*
Lot Number*
ID Number*
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Age*
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key816729
ManufacturerMOHAWK DENTAL SUPPLY CO., INC.
Manufacturer Address* SYRACUSE NY 13206 US


Patients

Patient NumberTreatmentOutcomeDate
101. Life Threatening; 2. Other 2007-03-16