TOWER WRIST JOINT 10100014

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a distributor report with the FDA on 2019-02-04 for TOWER WRIST JOINT 10100014 manufactured by Acumed Llc.

Event Text Entries

[134896041] Returned product examined visually under magnification. All threaded features were inspected and found to be in good condition. Welds connecting the arm sleeves to the locking joint have cracked. Additional mdrs associated with this event: 3025141-2019-00097: rosette knob 1, 3025141-2019-00098: rosette knob 2, 3025141-2019-00100: tower upper arc.
Patient Sequence No: 1, Text Type: N, H10


[134896042] During surgery on the patient's arm, it was supported by the arc wrist tower. During the surgery, the surgeon found tiny metal flakes in the wound. The flakes were coming from the arc wrist tower. All of the metal flakes were not successfully removed from the wound, as some were sticking to the synovium even when washed thoroughly. There was a delay in surgery as a result of this situation.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3025141-2019-00099
MDR Report Key8301468
Report SourceDISTRIBUTOR
Date Received2019-02-04
Date of Report2019-01-10
Date of Event2019-01-10
Date Mfgr Received2019-01-10
Device Manufacturer Date2018-01-15
Date Added to Maude2019-02-04
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. MICKI ANDERSON
Manufacturer Street5885 NE CORNELIUS PASS ROAD
Manufacturer CityHILLSBORO OR 97124
Manufacturer CountryUS
Manufacturer Postal97124
Manufacturer Phone8886279957
Manufacturer G1ACUMED LLC
Manufacturer Street5885 NE CORNELIUS PASS ROAD
Manufacturer CityHILLSBORO OR 97124
Manufacturer CountryUS
Manufacturer Postal Code97124
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTOWER WRIST JOINT
Generic NameAPPARATUS, TRACTION, NON-POWERED
Product CodeHST
Date Received2019-02-04
Returned To Mfg2019-01-14
Model Number10100014
Catalog Number10100014
Lot Number364026
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerACUMED LLC
Manufacturer Address5885 NE CORNELIUS PASS ROAD HILLSBORO OR 97124 US 97124


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2019-02-04

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.