ELECSYS PROLACTIN ASSAY 07027737190

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2019-02-04 for ELECSYS PROLACTIN ASSAY 07027737190 manufactured by Roche Diagnostics.

Event Text Entries

[134979293] The event occurred in (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[134979294] The customer complained of a questionable elecsys prolactin assay result for 1 patient tested on a cobas 8000 e 801 module compared to an architect analyzer. The initial prolactin result was 15. 1 ng/ml. The same patient sample was retested on (b)(6) 2019 with a prolactin result of 15. 3 ng/ml. On (b)(6) 2019 the same patient sample was tested on an architect with a prolactin result of 62. 8 ng/ml. The erroneous result was reported outside of the laboratory. There was no allegation of an adverse event. The cobas e801 serial number was requested but was not provided. The investigation is currently ongoing.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2019-00443
MDR Report Key8302077
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2019-02-04
Date of Report2019-03-27
Date of Event2019-01-08
Date Mfgr Received2019-01-21
Date Added to Maude2019-02-04
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameELECSYS PROLACTIN ASSAY
Generic NameRADIOIMMUNOASSAY, PROLACTIN (LACTOGEN)
Product CodeCFT
Date Received2019-02-04
Model NumberNA
Catalog Number07027737190
Lot NumberASKU
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2019-02-04

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.