API? RAPIDEC? CARBA NP 417824

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2019-02-05 for API? RAPIDEC? CARBA NP 417824 manufactured by Biomerieux, Sa.

Event Text Entries

[137022566] A customer in the united states contacted biom? Rieux to report a false negative result for a klebsiella pneumonia isolate in association with the rapidec? Carba np product. The customer stated they tested the klebsiella pneumonia isolate twice with microscan; the result indicated a possible carbapenem resistant enterobacteriaceae (cre). Testing with rapidec carba np provided a "negative" result. The result provided to the physician was "cre with carba np test negative". The patient isolate was sent to an external laboratory for further testing because the customer remained suspicious of cre. Pcr testing was performed = positive for oxa 48 carbapenemase. The report to physician was modified that pcr testing was cre positive. There is no indication or report from the laboratory or physician that the discrepant result led to any adverse event related to the patient's state of health. Biom? Rieux requested strain submittal from the customer. Biom? Rieux investigation will be conducted.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9615754-2019-00006
MDR Report Key8304917
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2019-02-05
Date of Report2019-05-10
Date Mfgr Received2019-04-15
Device Manufacturer Date2017-02-23
Date Added to Maude2019-02-05
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. JEFF SCANLAN
Manufacturer Street100 RODOLPHE STREET
Manufacturer CityDURHAM NC 27712
Manufacturer CountryUS
Manufacturer Postal27712
Manufacturer G1BIOMERIEUX, SA
Manufacturer Street3 ROUTE DE PORT MICHAUD
Manufacturer CityLA BALME, LES GROTTES ISERE 38390
Manufacturer CountryFR
Manufacturer Postal Code38390
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameAPI? RAPIDEC? CARBA NP
Generic NameRAPIDEC? CARBA NP
Product CodePTJ
Date Received2019-02-05
Catalog Number417824
Lot Number1005739620
Device Expiration Date2019-02-23
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBIOMERIEUX, SA
Manufacturer Address3 ROUTE DE PORT MICHAUD LA BALME, LES GROTTES ISERE 38390 FR 38390


Patients

Patient NumberTreatmentOutcomeDate
10 2019-02-05

© 2026 FDA.report