EZ-IO 25MM NEEDLE SET-EU CONFIGURATION 9001-EU-001

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2019-02-07 for EZ-IO 25MM NEEDLE SET-EU CONFIGURATION 9001-EU-001 manufactured by Teleflex Medical.

Event Text Entries

[135382420] (b)(4). The complaint sample is not available for return. Functional testing and investigation activities cannot be conducted. The complaint root cause cannot be established. The complaint cannot be confirmed. The device history review file is not available for review in the us. The ifu states, "attach luer-lock syringe to hub of catheter. Withdraw the catheter by applying traction while rotating the syringe and catheter clockwise. Maintain axial alignment during removal. Do not rock or bend the catheter. " a review of the certificate of compliance could not be accomplished as the product lot number was not reported. Teleflex will continue to monitor and trend related events.
Patient Sequence No: 1, Text Type: N, H10


[135382421] It was reported that a patient has had an io needle removed from the shin/tibia but the needle snapped and the needle fragments were retained in the shin/tibia. The patient will require an operation to remove the needle fragments when he is more stable.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3011137372-2019-00049
MDR Report Key8315761
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2019-02-07
Date of Report2018-12-21
Date of Event2018-12-21
Date Mfgr Received2018-12-21
Date Added to Maude2019-02-07
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationRISK MANAGER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactEFFIE JEFFERSON
Manufacturer Street3015 CARRINGTON MILL BLVD
Manufacturer CityMORRISVILLE 27560
Manufacturer CountryUS
Manufacturer Postal27560
Manufacturer Phone9194332672
Manufacturer G1TELEFLEX MEDICAL
Manufacturer Street3015 CARRINGTON MILL BLVD
Manufacturer CityMORRISVILLE NC 27560
Manufacturer CountryUS
Manufacturer Postal Code27560
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameEZ-IO 25MM NEEDLE SET-EU CONFIGURATION
Generic NameTUBING, PRESSURE AND ACCESSORI
Product CodeBYX
Date Received2019-02-07
Catalog Number9001-EU-001
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerTELEFLEX MEDICAL
Manufacturer AddressATHLONE


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2019-02-07