VIDAS? BRAHMS PROCALCITONIN 30450-01

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-02-07 for VIDAS? BRAHMS PROCALCITONIN 30450-01 manufactured by Biomerieux Sa.

Event Text Entries

[136343160] A customer in the united states contacted biom? Rieux to report the occurrence of a (b)(6) sample out of range (result was 1. 49ng/ml, acceptable range 1. 56 - 2. 30ng/ml) for the vidas brahms procalcitonin (pct) assay. Repeat testing obtained results of 1. 42 and 1. 30 ng/ml. There is no indication or report from the laboratory or physician that the discrepant result led to any adverse event related to the patient's state of health. There was no patient associated with the cap survey sample. Biom? Rieux investigation will be conducted.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number8020790-2019-00008
MDR Report Key8317462
Date Received2019-02-07
Date of Report2019-04-01
Date of Event2018-09-03
Date Mfgr Received2019-03-06
Device Manufacturer Date2018-04-26
Date Added to Maude2019-02-07
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. JEFF SCANLAN
Manufacturer Street100 RODOLPHE STREET
Manufacturer CityDURHAM NC 27712
Manufacturer CountryUS
Manufacturer Postal27712
Manufacturer G1BIOMERIEUX SA
Manufacturer StreetCHEMIN DE L ORME
Manufacturer CityMARCY L ETOILE, RHONE 69280,
Manufacturer CountryFR
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameVIDAS? BRAHMS PROCALCITONIN
Generic NameVIDAS? BRAHMS PCT
Product CodeNTM
Date Received2019-02-07
Catalog Number30450-01
Lot Number1006455600
Device Expiration Date2019-04-12
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBIOMERIEUX SA
Manufacturer AddressCHEMIN DE L ORME MARCY L ETOILE, RHONE 69280, FR


Patients

Patient NumberTreatmentOutcomeDate
10 2019-02-07

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.