DEFOGGER 28-0101B

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a distributor report with the FDA on 2019-02-07 for DEFOGGER 28-0101B manufactured by Deroyal Industries, Inc..

Event Text Entries

[136298168] Investigation summary: an internal complaint (b)(4) was received indicating that the x-ray element separated from the anti-fog foam pad (finished good 28-0101b) during a procedure. The x-ray detectable strip fell into the patient but was located after detaching. There was no patient injury. A sample was unavailable to return for evaluation. The complaint investigator reviewed the work order for the reported lot number (48028715). There were 27 defects from this lot; however, these parts were rejected due to issues with the paper backing, which is unrelated to the reported event. No discrepancies were identified in the work order that could have contributed to the reported failure. No change has been made to the raw materials or the converting process for the anti-fog pad. The investigation is ongoing at this time. When new and critical information is available, this report will be updated.
Patient Sequence No: 1, Text Type: N, H10


[136298169] The x-ray detectable strip came off the anti-fog pad during a procedure.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1060680-2019-00001
MDR Report Key8317853
Report SourceDISTRIBUTOR
Date Received2019-02-07
Date of Report2019-02-27
Date of Event2019-01-11
Date Mfgr Received2019-01-11
Date Added to Maude2019-02-07
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSARAH BENNETT
Manufacturer Street200 DEBUSK LANE
Manufacturer CityPOWELL TN 37849
Manufacturer CountryUS
Manufacturer Postal37849
Manufacturer Phone8653626112
Manufacturer G1DEROYAL INDUSTRIES, INC.
Manufacturer Street1703 HIGHWWAY 33 SOUTH
Manufacturer CityNEW TAZEWELL TN 37825
Manufacturer CountryUS
Manufacturer Postal Code37825
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameDEFOGGER
Generic NameANTI-FOG SOLUTION
Product CodeOCT
Date Received2019-02-07
Model Number28-0101B
Lot Number48028715
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerDEROYAL INDUSTRIES, INC.
Manufacturer Address185 RICHARDSON WAY MAYNARDVILLE TN 37807 US 37807


Patients

Patient NumberTreatmentOutcomeDate
10 2019-02-07

© 2026 FDA.report