Brachytherapy RPG-18

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-02-08 for Brachytherapy RPG-18 manufactured by Riverpoint Medical.

Event Text Entries

[137284516] Riverpoint medical received a complaint regarding holes in packaging for the rpg-18 brachytherapy grid product. Riverpoint performed a visual inspection of packaging on the products and an additional dye penetration test was conducted and confirmed breach of sterile barrier. There have been no reported adverse events associated with the packaging failure or reports of use of breached packaging product. A product removal was initiated by riverpoint on 08/february/2019 (initial removal report 3006981798-02072019-002-r) for the rpg-18 grid product.
Patient Sequence No: 1, Text Type: N, H10


[137284517] Riverpoint medical received a customer complaint on (b)(6) 2019 indicating a possible issue with the integrity of packaging for the rpg-18 grid product. According to the reporter, grid product was returned due to breached packaging.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3006981798-2019-00001
MDR Report Key8323805
Date Received2019-02-08
Date of Report2019-02-07
Date of Event2019-01-10
Device Manufacturer Date2018-11-12
Date Added to Maude2019-02-08
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. EDWIN ANDERSON
Manufacturer Street825 NE 25TH AVE.
Manufacturer CityPORTLAND OR 97232
Manufacturer CountryUS
Manufacturer Postal97232
Manufacturer Phone5035178001
Manufacturer G1RIVERPOINT MEDICAL
Manufacturer Street825 NE 25TH AVE.
Manufacturer CityPORTLAND OR 97232
Manufacturer CountryUS
Manufacturer Postal Code97232
Single Use3
Remedial ActionOT
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameBrachytherapy
Generic NameBRACHYTHERAPY GRID
Product CodeIWJ
Date Received2019-02-08
Model NumberRPG-18
Lot Number18101521
Device AvailabilityN
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerRIVERPOINT MEDICAL
Manufacturer Address825 NE 25TH AVE. PORTLAND OR 97232 US 97232


Patients

Patient NumberTreatmentOutcomeDate
10 2019-02-08

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.