ULTRA-CATCH NT (2.2 FR X 115 CM) STONE RETRIEVAL DEVICE NT4W22115

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2019-02-08 for ULTRA-CATCH NT (2.2 FR X 115 CM) STONE RETRIEVAL DEVICE NT4W22115 manufactured by Gyrus Acmi, Inc.

Event Text Entries

[138540164] The referenced device and original packaging was returned to olympus for evaluation. The evaluation confirmed the original (outer) box was severely crushed and opened as its seal was torn off. The (inner) packing tray containing the device inside is also crushed with multiple dents on the edge of the package, likely causing the packaging tray seal label to lift up and become slightly open; therefore, the sterile barrier was breached. The device appeared to be intact, and in new condition. Based on the evaluation, the reported event is likely due to improper handling from excessive force applied during transit or storage. The dhr review for this device was conducted. All records indicate that the device was manufactured and tested in accordance with all applicable procedures and met all final product release criteria.
Patient Sequence No: 1, Text Type: N, H10


[138540165] Olympus was informed that upon receipt of the device the device packaging was observed to be heavily damaged and the unit was unusable. There was no patient involvement. It was not specified if the sterile packaging of the device had been breached.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2951238-2019-00428
MDR Report Key8324279
Report SourceUSER FACILITY
Date Received2019-02-08
Date of Report2019-02-08
Date of Event2018-12-13
Date Mfgr Received2019-01-15
Date Added to Maude2019-02-08
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. CONNIE TUBERA
Manufacturer Street2400 RINGWOOD AVENUE
Manufacturer CitySAN JOSE CA 95131
Manufacturer CountryUS
Manufacturer Postal95131
Manufacturer Phone4089355124
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameULTRA-CATCH NT (2.2 FR X 115 CM) STONE RETRIEVAL DEVICE
Generic NameSTONE RETRIEVAL BASKET
Product CodeFGO
Date Received2019-02-08
Returned To Mfg2019-01-08
Model NumberNT4W22115
Lot NumberMK803710
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerGYRUS ACMI, INC
Manufacturer Address136 TURNPIKE ROAD SOUTHBOROUGH MA 01772 US 01772


Patients

Patient NumberTreatmentOutcomeDate
10 2019-02-08

© 2026 FDA.report