IMMULITE 2000 FREE PSA L2KPF2

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2019-02-11 for IMMULITE 2000 FREE PSA L2KPF2 manufactured by Siemens Healthcare Diagnostics Products, Limited.

Event Text Entries

[138556588] Assay adjustment (calibration) and quality control data from the customer has been provided for investigation. A siemens customer care center (ccc) specialist stated that the patient was not taking any medication or supplement. Arrangements are being made to send the sample to siemens for investigation. The cause of the discordant, falsely elevated results with the immulite 2000 free psa assay is unknown. Siemens is investigating the issue. Mdr 2432235-2019-0053 was filed for the same issue.
Patient Sequence No: 1, Text Type: N, H10


[138556589] Discordant, falsely elevated results were obtained with two reagent lots of the immulite 2000 free psa assay on an immulite 2000 xpi instrument. The discordant results were reported to the physician(s) and were questioned. The same samples were repeated on an alternate platform. The results from the alternate platform were lower and considered to be correct. Corrected reports were issued to the physician(s). There are no known reports of any impact to patient treatment or intervention due to the discordant, falsely elevated free psa results. There are no known reports of a delay in administering treatment or medical intervention to the patients due to the discordant, falsely elevated free psa results.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2432235-2019-00054
MDR Report Key8325244
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2019-02-11
Date of Report2019-03-21
Date of Event2018-12-10
Date Mfgr Received2019-03-01
Date Added to Maude2019-02-11
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactDEVYANI CHAUDHURI
Manufacturer Street511 BENEDICT AVENUE
Manufacturer CityTARRYTOWN NY 10591
Manufacturer CountryUS
Manufacturer Postal10591
Manufacturer Phone9145242637
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS, LIMITED
Manufacturer StreetUK REGISTRATION #: 3002806944 GLYN RHONWY
Manufacturer CityLLANBERIS, GWYNEDD LL554EL
Manufacturer CountryUK
Manufacturer Postal CodeLL55 4EL
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameIMMULITE 2000 FREE PSA
Generic NameIMMULITE 2000 FREE PSA
Product CodeMTG
Date Received2019-02-11
Model NumberIMMULITE 2000 FREE PSA
Catalog NumberL2KPF2
Lot Number311
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS, LIMITED
Manufacturer AddressGLYN RHONWY LLANBERIS, GWYNEDD LL554EL UK LL55 4EL


Patients

Patient NumberTreatmentOutcomeDate
10 2019-02-11

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.