ELECSYS VITAMIN D TOTAL II 07028148190

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2019-02-11 for ELECSYS VITAMIN D TOTAL II 07028148190 manufactured by Roche Diagnostics.

Event Text Entries

[135871910] This event occurred in (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[135871911] The customer questioned results for 15 patient samples tested for elecsys vitamin d total ii (vitamin d total ii) on a cobas e 801 module. Based on the data provided the results for 11 patient samples were discrepant. The initial results were reported outside of the laboratory. The customer thinks the issue is related to one particular reagent pack. The customer repeated patient samples on a cobas 8000 e 602 module using the vitamin d total assay. The customer also repeated the patient samples on the e801 module in question using the vitamin d total ii assay. The customer then repeated the patient samples using a new reagent pack on the e602 module. There was no allegation that an adverse event occurred. The e801 module serial number was (b)(4). The customer's maintenance was up to date. The measuring cell was last replaced in nov-2018. Liquid flow cleaning was last performed on 22-jan-2019.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2019-00559
MDR Report Key8327002
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2019-02-11
Date of Report2019-04-24
Date of Event2019-01-28
Date Mfgr Received2019-01-28
Date Added to Maude2019-02-11
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameELECSYS VITAMIN D TOTAL II
Generic NameVITAMIN D TEST SYSTEM
Product CodeMRG
Date Received2019-02-11
Model NumberNA
Catalog Number07028148190
Lot Number35568400
Device Expiration Date2019-08-31
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2019-02-11

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.